The CLD and the HCLD shall be calibrated in the most common
                           operating range, following the manufacturer's specifications using
                           zero and span gas (the NO content of which should amount to about
                           80% of the operating range and the NO2 concentration of
                           the gas mixture to less than 5% of the NO concentration). The NOx analyser
                           must be in the NO mode so that the span gas does not pass through
                           the converter. The indicated concentration shall be recorded.