5.21.1 The water flow rate of an open water mist
nozzle with its strainer or filter should be measured at its rated
working pressure. The nozzle and strainer or filter should then be
installed in test apparatus described in Figure 3 and subjected to 30
min of continuous flow at rated working pressure using contaminated
water which has been prepared in accordance with 5.21.3.
Figure 3 Clogging Test Apparatus
5.21.2 Immediately following the 30 min of continuous
flow with the contaminated water, the flow rate of the nozzle and
strainer or filter should be measured at rated working pressure. No
removal, cleaning or flushing of the nozzle, filter or strainer is
permitted during the test.
5.21.3 The water used during the 30 min of continuous
flow at rated working pressure specified in 5.21.1 should consist of 60 l of
tap water into which has been mixed 1.58 kg of contaminants which
sieve as described in Table
5. The solution should be continuously agitated during the
test.
Table 5 Contaminant for
Contaminated Water Cycling Test
Table 6 — Contaminant for the contaminated water cycling
test
|
SIEVE DESIGNATION
|
NOMINAL SIEVE OPENING, MM
|
GRAMS OF CONTAMINANT (+/- 5 PERCENT)
|
PIPE
SCALE
|
TOP SOIL
|
SAND
|
No. 25
|
0.706
|
—
|
456
|
200
|
No.
50
|
0.297
|
82
|
82
|
327
|
No.
100
|
0.150
|
84
|
6
|
89
|
No. 200
|
0.074
|
81
|
—
|
21
|
No. 325
|
0.043
|
153
|
—
|
3
|
|
Total
|
400
|
544
|
640
|
5.21.4 Alternative supply arrangements to the
apparatus shown in figure 3 may be used where damage to the pump is
possible. Restrictions to piping defined by note 2 of table 5 should
apply to such systems.