5.5.1.1 The analyser calibration curve shall be
                                 established by at least five calibration points (excluding zero) spaced
                                 as uniformly as possible. The highest nominal concentration shall
                                 be greater than or equal to 90% of full scale.
                              
                              
                              
                            
                           
                           
                           
                              
                              
                               
                                 5.5.1.2 The calibration curve is calculated by
                                 the method of least squares. If the resulting polynomial degree is
                                 greater than 3, the number of calibration points (zero included) shall
                                 be at least equal to this polynomial degree plus 2.
                              
                              
                              
                            
                           
                           
                           
                              
                              
                               
                                 5.5.1.3 The calibration curve shall not differ
                                 by more than ±2% from the nominal value of each calibration
                                 point and by more than ±1% of full scale at zero.
                              
                              
                              
                            
                           
                           
                           
                              
                              
                               
                                 5.5.1.4 From the calibration curve and the calibration
                                 points, it is possible to verify that the calibration has been carried
                                 out correctly. The different characteristic parameters of the analyser
                                 shall be indicated, particularly: