5.4.1 Each normally used operating range shall
                           be calibrated.
                        
                        
                        
                      
                     
                     
                     
                        
                        
                         
                           5.4.2 Using purified synthetic air (or nitrogen),
                           the CO, CO2, NOx and O2 analysers
                           shall be set at zero.
                           
                        
                        
                        
                      
                     
                     
                     
                        
                        
                         
                           5.4.3 The appropriate calibration gases shall
                           be introduced to the analysers, the value recorded, and the calibration
                           curve established according to 5.5 below.
                        
                        
                        
                      
                     
                     
                     
                        
                        
                         
                           5.4.4 The zero setting shall be rechecked and
                           the calibration procedure repeated, if necessary.