Water and CO2 may interfere with the CO analyser
                           performance. Therefore, a CO2 span gas having a concentration
                           of 80 to 100% of full scale of the maximum operating range used during
                           testing shall be bubbled through water at room temperature and the
                           analyser response recorded. The analyser shall not be more than 1%
                           of full scale for ranges greater than or equal to 300 ppm or more
                           than 3 ppm for ranges below 300 ppm.